Jenscare Scientific Co., Ltd. (Jenscare) has disclosed 12-month follow-up results from EU TRINITY, its prospective, multi-center, single-arm clinical trial evaluating the LuX-Valve Plus transcatheter tricuspid valve replacement system in patients with severe tricuspid regurgitation (TR) at elevated surgical risk.
The study enrolled 149 patients across 20 sites—18 of which are located in France, Germany, Spain, Denmark and the United Kingdom—featuring a mean age of 76.9 years and an average Tri-Score of 12.6. Baseline TR severity was high: 12.1% Severe, 51.0% Massive and 36.9% Torrential. More than 75% of patients required valve sizes of 55 mm to 70 mm, reflecting the product’s ability to accommodate annular dilation up to 70 mm.
Key procedural metrics and efficacy outcomes: • Device success rate: 96.6%. • Mean implantation time: 41.6 ± 19.6 minutes. • TR reduction: At 12 months, 80.6% of 126 echocardiographically evaluable patients presented with no more than Moderate TR; Severe TR persisted in only 2.4%, with no Massive or Torrential cases reported. • Functional gains: 92.1% of 127 evaluable patients achieved NYHA Class I/II at 12 months. • Quality-of-life improvement: Kansas City Cardiomyopathy Questionnaire scores rose by an average of about 17 points; mean 6-minute walk distance increased approximately 39 meters.
Safety profile (149 patients, 12-month adjudicated composite events): • Cardiovascular mortality: 6.0% (9 cases). • Stroke: 2.0% (3 cases). • Severe bleeding: 6.7% (10 cases). • New pacemaker implantation due to AV block: 8.7% (13 cases). • Composite event rate: 24.8%. No device-related pulmonary embolism or major access site complications were reported.
To date, LuX-Valve Plus has been implanted in over 1,200 patients worldwide, with follow-up extending beyond seven years in the earliest cases. The device holds EU MDR CE certification, has U.S. FDA approval for a Pivotal IDE trial with ongoing enrollment, and is eligible for CMS Category B Medicare Coverage.
These 12-month EU TRINITY findings reinforce the safety and efficacy profile of LuX-Valve Plus and expand its clinical evidence base, supporting Jenscare’s strategy to accelerate global commercialization across its tricuspid, mitral and aortic valve portfolios.